Instrument Functions
NRP2
116
User Manual 1173.9140.02 ─ 07
Fig. 4-46: Trace display mode
Plot
= channel / meas. function; display type
VBW / Avg = video bandwidth / averaging number
Trigger
= trigger slope and source
Offset / dB = level offset to take gain/attenuation at the sensor input into account
The following sections describe the functions of the control panel and diagram settings
in detail.
4.5.3.1
Controls in trace display mode
Display
Config Grid...
Opens the subdialog box for configuring the diagram, see
Pos Scale
Defines if the cursor keys position or scale the waveform.
"Pos"
The cursor keys shift the waveform in horizontal and vertical direction
"Scale"
The cursor keys expand and compress the waveform. When the hori-
zontal scaling is modified, the points (pixels) at the left of the screen
remain unchanged. When the vertical scaling is modified, the neutral
zone depends on the unit selected: the points in the center of the
screen remain unchanged when dBm, dB and dBµV are used. The
points at the bottom of the screen remain unchanged in all other dis-
plays.
Auto
Selects a vertical scaling that suits the waveform.
Trig...
Opens the "Trigger" dialog box. For description refer to
Filt...
Opens the "Filter" dialog box, see
chapter 4.5.3.3, "Filtering settings"
Displaying traces