Remote Control - Commands
NRP2
246
User Manual 1173.9140.02 ─ 07
This command corresponds to the <resolution> option of the high-level measurement
commands
CONFigure
,
FETCH
,
READ
and
MEASure
.
Note:
Automatic filter ranging must be switched on
(
SENSe<[1]..4>:AVERage:COUNt:AUTO ON
).
DISP<[1]..8>:RES
has an effect
on the setting
SENS<[1]..4>:AVER:COUN:AUTO:RES
of the sensors used in the
measurement function (
CALC<[1]..8>:MATH
). However, the resolution setting of the
sensors does not affect the resolution of values in the display.
Error messages:
-127
"Invalid numeric data"
: An invalid number was specified for the resolution.
Parameters:
<resolution>
1 | 0.1 | 0.01 | 0.001
lin = Linear scale of measurement result
log = Logarithmic scale of measurement result
1
lin = 1 significant place
log = No decimal place
0.1
lin = 2 significant places
log = 1 decimal place
0.01
lin = 3 significant places
log = 2 decimal places
0.001
lin = 4 significant places
log = 3 decimal places
*RST:
0.01
Example:
DISP:WIND1:RES 0.1
With a measured value of 123.456789 W, the first two places
are free of noise and thus 120 W is displayed.
DISPlay[:WINDow]:SELect
<select_window>
Selects a window.
Parameters:
<select_window>
1 | 2 | 3 | 4
*RST:
1
DISPlay[:WINDow]:SIZE
<window_size>
Sets the window size.
DISPlay