
Remote Control - Commands
NRP2
332
User Manual 1173.9140.02 ─ 07
TRIGger<[1]..4>:DELay:AUTO
<state>
TRIGger:ALL:DELay:AUTO
<state>
TRIG:DEL:AUTO ON
makes certain that a measurement is not started until the sensor
has settled. To this end, the delay value is automatically determined. This is especially
important when thermal sensors are used. The automatically determined delay is
ignored if a longer period is set with
TRIG<[1]..4>]:DEL
.
Parameters:
<state>
ON | OFF
*RST:
depending on sensor
TRIGger<[1]..4>:DTIMe
<dtime>
TRIGger:ALL:DTIMe
<dtime>
If the trigger level has been underrun with internal triggering, it is sometimes useful to
wait some time (drop-out time) before retriggering. The measurement of unwanted sig-
nal components can thus be suppressed. This setting parameter is used with pulsed
GSM signals, for example, if only specific timeslots are to be measured.
The figure below shows the importance of the drop-out time. If the timeslots BF and
DH are not to be measured, the drop-out time is set to a higher value than the period
EB (GD). In this case, the trigger events B and D are ignored.
Trigger level
Power
Time
Drop-Out
Time
Trigger hysteresis
A
B
C
D
E
F
G
H
Drop-Out
Time
Fig. 6-20: Significance of the drop-out time parameter
A similar response is achieved with the
SENSe:BURSt:DTOLerance
command in the
Burst mode. However, the Burst mode makes certain that a burst measurement is not
discontinued too early, whereas
TRIGger:DTIMe
suppresses a new measurement
that is performed too early.
Parameters:
<dtime>
float_value
*RST:
depending on sensor
TRIGger
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com