Remote Control - Commands
NRP2
215
User Manual 1173.9140.02 ─ 07
<source_list>
Value range: (
@
n) | (
@
n),(
@
m)
Default value: (
@
1) or (
@
1),(
@
2)
n and m may assume the values 1, 2, 3 or 4.
With this list, sensors are assigned to the primary and possibly to the secondary
channel of a calculate block. The number for each sensor is preceded by the
character
@
and the entire expression is enclosed in parentheses.
Examples:
(
@
1)is used for sensor
A.
The values (
@
3),(
@
2) are set for <source_list>. Sensor C is assigned to the
primary channel and sensor
B to the secondary channel. The calculate blocks
process the primary channel together with the secondary depending on the
desired function. If
:RATio
is selected as the calculation function, the result
from the calculate block is the ratio of the values measured by sensor
C and
sensor
B.
<start_exclude>
Value range: depending on sensor
Default value: depending on sensor
Unit: S
Amount of time at the beginning of a timeslot or integration period that should
not be taken into account. Corresponds to command:
[SENSe<[1]..4>]:TIMing:EXCLude:STARt
<statistics_size>
Value range: depending on sensor.
Default value: depending on sensor.
Unit: -
The number of test points on the time axis. Corresponds to command:
[SENSe<[1]..4>]:STATistics:SCALe:X:POINts
<tslot_width>
Unit: S
Value range: depending on sensor
*RST value: depending on sensor
Width of a timeslot in the Timeslot mode. Corresponds to command:
[SENSe<[1]..4>][:POWer]:TSLot:[AVG]:WIDTh
6.4.5 Calculation functions - High level commands
For high level commands, the calculation function is selected by the keyword in the
command header.
For high level commands, the calculation function is selected by the keyword in the
command header. The following calculation functions are available for calculating the
primary channel (PC) and the secondary channel (SC). The primary and secondary
channel are set by the parameter <source_list> in the parameter list.
The reference value for the calculation of relative values has to be set beforehand
using one of the following commands:
CALCulate[1..8]:RELative[:MAGNitude]
CALCulate[1..8]:RELative:AUTO
.
High-Level Measurement Commands (CONFigure, MEASure, READ and FETCh)
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com