Instrument Functions
NRP2
107
User Manual 1173.9140.02 ─ 07
Fig. 4-40: Setting the Unit
Only the units which are compatible with the selected measurement function are provi-
ded for selection. I.e. absolute power values are displayed in W, dBm or dB
μ
V , and
relative values in 1, dB or
Δ
%, respectively.
The following tables show how the R&S
NRP2 calculates the measurement results for
the different units either in absolute notation or relative to a reference value, see
ter 4.4.4, "Relative measurement"
Table 4-3: Formulas for absolute single channel power measurement
Function
Unit
W
dBm
dB
μ
V
Single
P
mW
1
lg
10
P
V
1
lg
20
Z
P
Table 4-4: Formulas for power ratio measurement
Function
Unit
W
dBm
dB
μ
V
Ratio (A,B)
B
A
P
P
lg
10
1
100
B
A
P
P
B
A
P
P
SWR (A,B)
–
–
A
B
A
B
P
P
P
P
1
1
RCoeff (A,B)
–
–
A
B
P
P
RLoss (A,B)
B
A
P
P
lg
10
–
–
Configuring measurements