Annex
NRP2
383
User Manual 1173.9140.02 ─ 07
Example:
MEAS:ARRay? (1000)
1000 measured values are recorded in the ContAv mode. After termination of the last
measurement, the results are stored in the output queue. There is no triggering as in
the ContAv mode. The measurement result is a list of 1000 measured values separa-
ted by a comma.
The keywords for processing two sensors can be added to the
MEAS:TSL
,
MEAS:BURS
and
MEAS:ARRay
commands, but the
:SWR
,
:RLOS
and
:REFL
functions are not avail-
able.
Example:
MEAS:TSL:RAT? 577 us, 8, 18 us, 18 us, DEF, 3, (@1), (@3)
The above timeslot example is extended to two sensors. The measurement result is a
list of 8 measured values separated by a comma for the 8 timeslots; these values were
obtained by division from the measured values of sensors 1 and 3.
A.4.1.5
Selecting the output unit for measured values
So far, no information has been given on the unit of the measured values. It is possible
to modify the output unit of the measured values. A distinction is made as to whether
the result represents a power or a power ratio. Relative measurements always deliver
power ratios. This also applies to the
:RATio
,
:SWR
,
:RLOS
and
:REFL
functions. The
units are set by means of the following commands:
UNIT:POWer DBM | W | DBUV
and
UNIT:RATio DB | DPCT | O
The default settings are
DBM
for power values and
DB
for power ratios.
Physical
unit
SCPI
notation
Meaning
dBm
DBM
Power in dB referenced to 1 mW: x / W = 10 log ( x / (1 mW) ) / dBm
Watt
W
dBµV
DBUV
Power in dB referred to (1µV)
2
/ 50 (: x dBm ( (x + 107) dBµV
dB
DB
Δ
%
DPCT
Deviation from 100% in %: x (% =
Δ
x + 100) %
1
O
When the
:SWR
,
:RLOS
and
:REFL
functions are used, the output unit, however, is
implicitly set by the
MEAS
command:
Programming Measurements