Remote Control Fundamentals
NRP2
184
User Manual 1173.9140.02 ─ 07
5.8.4 Status Reporting System
The status reporting system collects information on the device status and makes it
available to the output unit on request. Structure and function of this system are descri-
bed in the following section.
5.8.5 Output Unit
The data output unit collects the information requested by the controller from the data
set management. It processes the information in line with the SCPI rules and makes it
available in the output buffer.
5.8.6 Command Sequence and Command Synchronization
All commands are executed in the order in which they are sent. There is no overlap-
ping of command processing. (Exception: the commands
INITiate
and
CALibrate:ZERO:AUTO ONCE
). The user therefore can determine the sequence
of execution.
If a long program message is terminated by
*OPC
or
*OPC?
, the end of command pro-
cessing is signalled to the controller, provided the controller has been adequately pro-
grammed (see following
Synchronization with *OPC, *OPC? and *WAI
Table 5-2: Synchronization with *OPC, *OPC? and *WAI
Command
Effect
Programming of controller
*OPC
Sets the Operation Complete bit in the ESR
after all commands sent prior to *OPC have
been processed.
Set bit 0 in the ESE of the R&S
NRP2.
Set bit 5 in the SRE of the R&S
NRP2.
Wait for service request (SRQ).
*OPC?
Writes 1 in the output buffer after all com-
mands sent prior to
*OPC?
have been pro-
cessed.
IEC/IEEE bus: the R&S
NRP2 is addressed
as a talker.
*WAI
Commands can be sent but they are only
executed after completion of all commands
sent prior to
*WAI
.
Dividing commands timewise before/after
*WAI
serializes the command sequence.
*WAI
is sent directly after the command
that should be completed before other com-
mands can be executed.
5.9 Automatic Settings for Manual Selection of a Mea-
surement Mode
When the measurement mode of a sensor is manually selected ("Sensor"
→
"Mode"),
automatic settings are made for the trigger system and timing, which cannot be
changed by manual control. This considerably simplifies operation. These settings are
not changed when the measurement mode is selected by remote control (
SENSe:
FUNCtion
) and have to be set to the desired value by appropriate SCPI commands.
Automatic Settings for Manual Selection of a Measurement Mode