Getting Started
NRP2
48
User Manual 1173.9140.02 ─ 07
The parameter "Dropout" helps to reliably detect the end of modulated-signal
bursts (e.g. NADC). With unmodulated pulses as in this case, it should be set to 0.
Fig. 2-39: Burst av result window
For a detailed description of burst av power measurement, see
suring the average power of burst signals"
The power of burst can be measured also by means of timegates (see
"Measuring the average power in gates"
on page 71 and
on page 120). In contrast to the burst mode which automatically detects
the burst end, the length and position of the gate over time must be exactly defined.
Measuring average burst power (Burst Av mode)