Getting Started
NRP2
45
User Manual 1173.9140.02 ─ 07
Note:
If you do not see a trace, the trigger threshold probably does not match the
signal. In this case, a highlighted appears in the info panel of the trace window.
Continue with the next step to set the trigger threshold.
Further measurement modes can be selected via the "Display" softkey.
Based upon the trace display, gates, timeslots, and markers can be added for further
analysis. Automatic pulse measurements can be carried out and a statistic analysis of
the power envelope can be performed.
Settings for reliable triggering
1. Press the "Trigger" softkey in the trace display.
The "Trigger" dialog box opens.
Fig. 2-34: Trigger dialog box, page 1
A dotted line indicates the trigger threshold in the diagram.
Fig. 2-35: Trace display / trigger threshold
2. Select the "Normal" trigger mode.
3. Select internal triggering ("Source" = "Internal")
4. Set the positive trigger slope.
5. Set the trigger threshold in the "Level" entry field.
Tip:
A
in the info panel shows, that the sensor doesn’t trigger. In this case
use the DEL/1TRIG hardkey in order to manually trigger the sensor once, and set
the trigger settings accordingly.
Graphically representing power versus time