Remote Control - Commands
NRP2
287
User Manual 1173.9140.02 ─ 07
[SENSe<[1]..4>]:TRACe:AVERage:TCONtrol
<terminal_control>
As soon as a new single value is determined, the filter window is advanced by one
value so that the new value is taken into account by the filter and the oldest value is
forgotten.
[SENSe<[1]..4>]:AVERage:TCONtrol
(terminal control) then deter-
mines in the Trace mode whether a new result will be calculated immediately after a
new measured value is available (
MOVing
) or only after an entire range of new values
is available for the filter (
REPeat
).
Parameters:
<terminal_control>
MOVing | REPeat
*RST:
REP; (PRES) and SYST:PRES:MOV;
[SENSe<[1]..4>]:TRACe:MEASurement:ALGorithm
<algorithm>
Selects the analysis algorithm for detecting the pulse top and the pulse base power of
a pulsed signal. These two power levels are fundamental for further signal analysis.
Parameters:
<algorithm>
HISTogram
Computes the pulse levels by analyzing the histogram of the
trace data. The pulse top level is taken on the mean value of all
points representing the pulse top. Similarly the pulse base is
determined from the points representing the pulse base.
Note: This algorithm is recommended for analyzing most of the
pulse signals.
INTegration
Computes the pulse as a reference top power by fitting a rectan-
gle pulse of same energy into the pulse signal.
Note: Use this algorithm for pulse signals with modulation or
when the pulse energy must be taken into account, e.g. when
comparing the measurement result with a dual thermal power
sensor.
PEAK
Assigns the pulse peak power to the pulse top level.
*RST:
depending on the sensor
[SENSe<[1]..4>]:TRACe:MEASurement:AUTO[:STATe]
<state>
Switches on/off automatic transfer of the measured pulse parameters after each trace.
This enables a synchronous display of the trace and the pulse data during a continu-
ous measurement with
INIT:CONT ON
.
For pulse measurements, this settings has to be switched
ON
. See also
.
Parameters:
<state>
ON | OFF
*RST:
depending on the sensor
SENSe (Sensor Commands)