Remote Control - Commands
NRP2
255
User Manual 1173.9140.02 ─ 07
The following commands affect this table:
●
MEMory:CLEar[:NAMe] <string>
●
MEMory:CLEar:TABLe
●
all
MEMory:TABLe
commands
The selection is not changed in case of an
*RST
or a
SYSTem:PRESet
RST value:
none
Error messages:
-256
"File name not found"
: A table named <select> is not known .
6.10 OUTPut
The OUTPut system serves for generator configuration. The R&S
NRP2 can be equip-
ped with an optional RF test generator which produces a reference power of 1
mW at
50
MHz. Moreover, the analog outputs at the rear of the power meter can be config-
ured.
.....................................................................................255
OUTPut:RECorder<[1]..2>:LIMit:LOWer
........................................................................... 256
OUTPut:RECorder<[1]..2>:LIMit:LOWer:POWer
................................................................256
OUTPut:RECorder<[1]..2>:LIMit:LOWer:RATio
................................................................. 256
OUTPut:RECorder<[1]..2>:LIMit:UPPer
............................................................................257
OUTPut:RECorder<[1]..2>:LIMit:UPPer:POWer
................................................................ 257
OUTPut:RECorder<[1]..2>:LIMit:UPPer:RATio
.................................................................. 257
OUTPut:RECorder<[1]..2>[:STATe]
................................................................................. 258
..........................................................................................258
.............................................................................................. 260
OUTPut:RECorder<[1]..2>:FEED
<calculate_block>
Assigns a calculate block to one of the two analog outputs. On the basis of the set
characteristic, the measured value is converted to a DC voltage that can be tapped at
the appropriate output.
Parameters:
<calculate_block>
"CALC1", "CALC2", "CALC3", "CALC4"
*RST:
output 1: "CALC1"; output 2: "CALC2"
OUTPut