[SENSe<[1]..4>]:STATistics:SCALe:X:POINts
<number>
Determines the number of samples on the power axis for the Statistics modes or
returns the current setting. Due to the limited storage capacity of the R&S
NRP2, the
number is limited to 1024.
Parameters:
<number>
int_value
Range:
1 ... 1024
[SENSe<[1]..4>]:STATistics:SCALe:X:RANGe
<range>
Defines, together with the
SENS:STAT:SCAL:X:RLEV
command, the range on the
power axis over which the statistics function (CCDF/PDF) is calculated (see
Setting the power axis for statistics measurements
).
Parameters:
<range>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: DB
[SENSe<[1]..4>]:STATistics:SCALe:X:RLEVel
<rlevel>
Defines, together with the
SENS:STAT:SCAL:X:RANG
command, the range on the
power axis over which the statistics function (CCDF/PDF) is calculated (see figure
below).
Power
Density/Distribution
SENS:STAT:SCAL:X:RANG
SENS:STAT:SCAL:X:RLEV
Fig. 6-14: Setting the power axis for statistics measurements
Unit:
DBM | W | DBUV