Instrument Functions
NRP2
140
User Manual 1173.9140.02 ─ 07
Statistics evaluation function
Note:
A linear (lin) or logarithmic (log) scale is available for "CCDF" and "CDF". The
"PDF" can only be displayed linearly.
"CCDF"
C
omplementary
C
umulative
D
istribution
F
unction. Displays the prob-
ability of the envelope power being higher than the corresponding X
axis power value.
"CDF"
C
umulative
D
istribution
F
unction.
Displays the probability of the envelope power being lower than the
corresponding X axis power value.
"PDF"
P
robability
D
ensity
F
unction.
Displays a normalized distribution density of the envelope power. The
measurement result is dimensionless and – in contrast to a non-nor-
malized "PDF" – is independent of the magnitude of the average
power value (Av). Integrating the normalized "PDF" over power pro-
vides the average power value (Av).
Remote command:
Absolute X-Scaling
Scales the X axis in terms of absolute power (dBm) if activated
Normalizes the X axis with respect to average power (dB) if deactivated.
Trace 1 / 2
Selects tab "Trace 1" or "Trace 2" to configure the statistic measurement for the 1st or
2nd trace.
An internal AWGN source is available for the second trace, thus providing a standard
signal reference.
Channel
Selects the channel for the respective trace. A statistical evaluation of combined
results from different channels, e.g. power ratios, is not supported.
Aperture
Sets the length of the aquisition interval with unsynchronized acquisition ("Cont", see
In addition to, and independent of, the setting of the aperture time for the measurement
of an average power value (Av) (
chapter 4.2.1.2, "Continuous average-power mea-
on page 66 ), you can use this parameter to set the length of the
unsynchronized time interval for the statistical analysis of the signal.
Remote command:
[SENSe<[1]..4>]:STATistics:TIME
Samples
Enters the required minimum number of samples.
Displaying traces