Getting Started
NRP2
28
User Manual 1173.9140.02 ─ 07
Zeroing
Zeroing should be performed whenever necessary. As a rule of thumb, this is the case
for measurements of power values less than a 1000 fold of the zero offset and after the
sensor has reached its operating temperature.
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
To zero the meter
1. Connect the sensor to a signal source
2. Switch off the power
3. Press the ZERO hardkey.
The "Zero" dialog box is displayed.
Fig. 2-2: Zeroing dialog box
4. Press the ZERO hardkey again to perform zeroing of all connected sensor chan-
nels ("Zero (All)") or press the appropriate softkey to select a specific sensor for
zeroing.
Zeroing of all connected sensors is performed.
The correction measurement takes several seconds. When completed, the
R&S
NRP2 displays a message indicating success or failure of zeroing.
Preset and zeroing