[SENSe<[1]..4>]:TRACe:MEASurement:DEFine:DURation:REFerence
<ref_level>
Sets the medial reference level in terms of percentage of the pulse power amplitude.
This level is used to define the pulse width, pulse start time and pulse stop time.
Unit: PCT | O
Parameters:
<ref_level>
float_value
Range:
depending on the sensor
*RST:
depending on the sensor
Default unit: PCT
[SENSe<[1]..4>]:TRACe:MEASurement:DEFine:TRANsition:HREFerence
<href_level>
Sets the high reference level in terms of percentage of the pulse power amplitude. The
"High Ref. Power" defines the end of the rising edge and the start of the falling edge of
the pulse. These values are needed for measurement of the rise / fall time.
Unit: PCT | O
Parameters:
<ref_level>
float_value
Range:
depending on the sensor
*RST:
depending on the sensor
Default unit: PCT
[SENSe<[1]..4>]:TRACe:MEASurement:DEFine:TRANsition:LREFerence
<lref_level>
Sets the low reference level in terms of percentage of the pulse power amplitude. The
"Low Ref. Power" defines the start of the rising edge and the end of the falling edge of
the pulse. These values are needed for measurement of the rise / fall time.
Unit: PCT | O
Parameters:
<ref_level>
float_value
Range:
depending on the sensor
*RST:
depending on the sensor
Default unit: PCT
[SENSe<[1]..4>]:TRACe:MEASurement:POWer:AVG?
Returns the measured average power in the current trace.
Unit: WATT
Usage:
Query only