Instrument Functions
NRP2
108
User Manual 1173.9140.02 ─ 07
Table 4-5: Formulas for sum and difference of power
Function
Unit
W
dBm
db
μ
V
Sum (A+B)
B
A
P
P
mW
1
lg
10
B
A
P
P
V
μ
1
lg
20
Z
P
P
B
A
Diff (A-B)
B
A
P
P
mW
1
lg
10
B
A
P
P
V
μ
1
lg
20
Z
P
P
B
A
Table 4-6: Formulas for relative measurement
Function
Unit
1
dB
Δ
%
Single
R
P
P
R
P
P
lg
10
1
100
R
P
P
Sum
Diff
R
B
A
P
P
P
mW
1
lg
10
B
A
P
P
1
100
R
B
A
P
P
P
Ratio
R
Y
X
B
A
P
P
P
P
R
Y
X
B
A
P
P
P
P
lg
10
1
100
R
Y
X
B
A
P
P
P
P
Table 4-7: Legend for formula tables
Variable
Meaning
P
P
A
P
B
P
R
(P
X
/ P
Y
)
R
Z
Power
Power in channel A (incident power)
Power in channel B (reflected power)
Reference (power)
Reference (power ratio)
Characteristic impedance, i.e. 50
Ω
Unit
The "Unit" drop-down list is used to set the result unit. Units that are provided for the
selected measurement function:
"W"
Single, Diff, Sum
"dBm | dB
μ
V"
Single, Diff, Sum
"
Δ
% | dB | 1"
Ratio
Configuring measurements
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com