Annex
NRP2
390
User Manual 1173.9140.02 ─ 07
Example
*RST
TRIG:SOUR INT
'triggering in response to a signal edge
TRIG:DEL 50 US
'starts measurement 50 µs after the trigger event
TRIG:HOLD 800 US
'starts next trigger event 800 µs
'after the trigger event at the earliest
INIT
FETCh?
Another way to modify the trigger system timing is to set a holdoff time (
TRIG:HOLD
).
After a trigger event, all following trigger events are ignored if they occur within the
holdoff time. This is to prevent signal components that have not yet decayed from cau-
sin triggering by mistake after the measurement has been completed. This makes
sense especially with internal triggering ("BurstAv" mode).
Ignored
trigger
event
Power
Time
TRIG:HOLDoff
TRIG:DELay
Trigger
event
Measurement
Trigger level
Fig. 1-7: Meaning of settings for TRIG:HOLD and TRIG:DEL
Exclude and Fence (Mid)
In the measurement modes Burst, Timeslot and Timegate, it is also possible to exclude
time domains from the integration of the measured value at the beginning and the end
of the measurement (
SENS:TIM:EXCL:STARt
and
SENS:TIM:EXCL:STOP
):
Example
*RST
SENS:FUNC "POW:BURS:AVG"
SENS:TIM:EXCL:STAR 10 US
'ignores 10 µs at the beginning of the measurement
SENS:TIM:EXCL:STOP 10 US
'ignores 10 µs at the end of the measurement
INIT
FETCh?
Programming Measurements