Remote Control - Commands
NRP2
257
User Manual 1173.9140.02 ─ 07
Parameters:
<lower_limit>
float_value
Range:
-200 DB .. 200 DB
*RST:
0 DB
Default unit: DB
OUTPut:RECorder<[1]..2>:LIMit:UPPer
<upper_limit>
Specifies the upper power limit of the characteristic for one of the two analog outputs
(see
Characteristic of analog outputs
). The unit corresponds to the output
unit of the measured value.
Value range:
Value ranges, depending on the unit
.
Unit:
Output unit of measured value of calculate block
.
Parameters:
<upper_limit>
float_value
*RST:
20 DBM or 10 DB
OUTPut:RECorder<[1]..2>:LIMit:UPPer:POWer
<upper_limit>
Specifies the upper power limit of the characteristic for one of the two analog outputs if
the associated calculate block returns a power (unit
DBM
,
W
or
DBUV
) as the measured
Characteristic of analog outputs
Unit:
DBM | W | DBUV
Parameters:
<upper_limit>
float_value
Range:
-120 DBM to 90 DBM
*RST:
20 DBM
Default unit: DBM
OUTPut:RECorder<[1]..2>:LIMit:UPPer:RATio
<upper_limit>
Specifies the upper power limit of the characteristic for one of the two analog outputs if
the associated calculate block returns a power ratio (unit
DB
,
DPCT
or
O
) as the mea-
Characteristic of analog outputs
Unit:
DB | DPCT | O
Parameters:
<upper_limit>
float_value
Range:
-200 DB .. 200 DB
*RST:
10 DB
Default unit: DB