Remote Control - Commands
NRP2
242
User Manual 1173.9140.02 ─ 07
Parameters:
<avalue>
NONE | NORMal | EXTRemes | STATistics
NONE
No additional information about the measured values is dis-
played.
NORMal
The display is a function of the
SENS:FUNCtion
of the mea-
surement mode of the primary channel and the
CALC:FEED
set-
ting.
Primary channel in the trace mode:
The top line displays the frequency of the primary channel. The
two lower lines and the main measured value window display
Average, Peak and PeakToAverage.
CALC:FEED
determines
which of the three quantities is displayed in the main measured
value window. The other two quantities are displayed on the
right-hand side, below the frequency.
Primary channel not in the trace mode:
Only the frequency of the primary channel and, if applicable, the
frequency of the secondary channel are displayed.
EXTReme
Minimum, maximum and peak-to-peak (maximum to minimum)
of all measured values displayed since the last
CALC:EXTR:RES
, PowerOn or AuxVal Reset are shown (man-
ual operation). With logarithmic units, the peak-to-peak value
equals the quotient of the measured values converted into linear
units.
STATistic
The mean value, the standard deviation and the number of all
measured values displayed since the last
CALC:EXTR:RES
,
PowerOn or AuxVal Reset are shown (manual operation) (see
also
).
*RST:
NONE
DISPlay[:WINDow]<[1]..4>:FORMat
<display_mode>
Selects the display mode for measured data.