Annex
NRP2
393
User Manual 1173.9140.02 ─ 07
*RST
SENS:FUNC "POW:AVG"
SENS:APER 10 MS
INIT
FETCh?
Timeslot mode:
SENS:FUNC "POW:TSL:AVG"
After the trigger event, the power is measured in timeslots, the number of which is
defined with
SENS:TSL:COUN
. The length of the timeslots is set with
SENS:TSL:WIDT
.
*RST
SENS:FUNC "POW:TSL:AVG"
SENS:TSL:WIDT 577 US
SENS:TSL:COUN 8
TRIG:SOUR EXT
INIT
FETCh?
BurstAv mode:
SENS:FUNC "POW:BURS:AVG"
Measurement of power pulses (bursts) with automatic detection of the beginning and
end of pulses. The trigger source setting is ignored. As described above, the conditions
for the detection of the pulse end can be set with commands
SENS:BURS:DTOL
and
TRIG:HYST
.
*RST
SENS:FUNC "POW:BURS:AVG"
SENS:BURS:DTOL 100 MS
TRIG:HYST 3 DB
INIT
FETCh?
Trace mode:
SENS:FUNC "XTIM:POW"
This mode is used to determine a large number of measured values in succession. The
number of required measured values is set with "SENS:SWE:POI" and the total mea-
surement time with
SENS:SWE:TIM
. Note that each measurement must be triggered
separately. This point is a difference between the "Trace" mode and the "Timeslot"
mode, in which it is only necessary to start the measurement for measuring the power
in all timeslots.
*RST
SENS:FUNC "XTIM:POW"
SENS:SWE:POIN 256
SENS:SWE:TIM 600 US
TRIG:COUN 256
'change to IDLE state after the 256th measurement
INIT
FETCh?
Programming Measurements