Remote Control - Commands
NRP2
336
User Manual 1173.9140.02 ─ 07
UNIT<[1]..8>:POWer
<output_unit>
Selects the output unit for the measured power values.
This setting also determines the unit for the parameters of the following commands:
●
CALC:REL
●
CALC:LIM:UPP
●
CALC:LIM:LOW
●
DISP:MET:UPP
●
DISP:MET:LOW
●
OUTP:REC:LIM:LOW
●
OUTP:REC:LIM:UPP
Parameters:
<output_unit>
DBM | W | DBUV
*RST:
DBM
UNIT<[1]..8>:POWer:RATio
<output_unit>
Selects the output unit for the measured power ratio values.
This setting also determines the unit for the parameters of the following commands:
●
CALC:REL
●
CALC:LIM:UPP
●
CALC:LIM:LOW
●
DISP:MET:UPP
●
DISP:MET:LOW
●
OUTP:REC:LIM:LOW
●
OUTP:REC:LIM:UPP
Error messages:
If the compute function of the associated calculate block is
:SWR
,
:RLOSs
or
:
REFLection
, then the unit cannot be changed and error message
26,"State not sup-
ported"
, is output.
Parameters:
<output_unit>
DB | DPCT | O
*RST:
DB
UNIT