Remote Control - Commands
NRP2
230
User Manual 1173.9140.02 ─ 07
CALCulate<[1]..8>:LIMit:UPPer[:DATA]
<upper_limit>
Sets an upper limit for the measured values. The R&S
NRP2 can respond when this
limit is exceeded (see other
CALCulate<[1]..8>:LIMit
commands).
Parameters:
<upper_limit>
float_value
The value range depends on the current output unit of the calcu-
late block's measured value, as shown in
.
The parameter unit is determined according to
unit of measured value of calculate block
*RST:
0 DBM or 0 DB
Default unit: see table referred to above.
CALCulate<[1]..8>:LIMit:UPPer[:DATA]:POWer
<power>
Sets a upper limit for the measured power. The R&S
NRP2 can respond when this limit
is exceeded (see other
CALCulate[1]..8>:LIMit
commands).
Parameters:
<power>
float_value; units: DBM | W | DBUV
Range:
-120 DBM to 90 DBM
*RST:
0 DBM
Default unit: DBM
CALCulate<[1]..8>:LIMit:UPPer[:DATA]:RATio
<ratio>
Sets an upper limit for the measured power ratios. The R&S
NRP2 can respond when
this limit is exceeded (see other
CALCulate<[1]..8>:LIMit
commands).
Parameters:
<ratio>
float_value, units: DB | DPCT | O
Range:
-200 DB to 200 DB
*RST:
0 DB
Default unit: DB
CALCulate<[1]..8>:LIMit:UPPer:STATe
<state>
Switches the monitoring function for the upper limit on or off.
Parameters:
<state>
ON | OFF
*RST:
OFF
Usage:
Setting only
CALCulate (Configuration of Calculate Blocks)