Remote Control - Commands
NRP2
271
User Manual 1173.9140.02 ─ 07
Parameters:
<mode>
string
"POWer:AVG"
ContAv
After a trigger event, the power is integrated over a time interval
(averaging) set with
SENS:POW:APER
.
"POWer:TSLot:AVG"
Timeslot
The power is simultaneously measured in up to 26 time win-
dows. The number of time windows is set with
SENS:POW:TSL:COUN
. The length of a time window is deter-
mined via
SENS:POW:TSL:WIDT
. The measurement result is
represented by a vector that can contain up to 26 indices and
contains the power of a time window at each index.
"POWer:TGATe:AVG"
Timegate
The Timegate mode is a pseudo-mode. It is used to determine
accurate measured values in the gates defined with the aid of
the Trace mode in manual operation. If one of the four gates is
selected (
SENS:TRAC:SEL
), the sensor is internally put into the
Timeslot mode and configured such that it determines the aver-
age power in this gate with high accuracy.
"POWer:BURSt:AVG"
Burst
In remote control, this measurement mode is very similar to the
ContAv mode. The integration time is, however, not predefined
but determined by the sensor with the aid of a burst detector.
The start of a burst is detected when the measurement signal
rises above the set trigger level. The end is set when the signal
drops below the trigger threshold.
SENS:POW:BURS:DTOL
defines the time interval during which a signal drop below the
trigger level is not interpreted as the end of the burst. In the
Burst mode, the set trigger source is ignored and
TRIG:SOUR
INT
is implicitly assumed.
"XPOWer:CCDF"
CCDF
The power in a settable time interval is measured and subse-
quently subjected to a statistical evaluation (CCDF). The mea-
surement time is set using the
SENS:STAT
commands. The
R&S
NRP2 returns in response to
READ?
or
FETCh?
an array of
density values which are spread over the power axis. The value
range of the power axis is also set using the
SENS:STAT:SCAL:X
commands.
CDF
The CDF does not represent a separate mode in the sensor. It is
calculated in the R&S
NRP2 by way of the relation
CDF = 1-CCDF.
"XPOWer:PDF"