Instrument Functions
NRP2
105
User Manual 1173.9140.02 ─ 07
To configure the measurement of
Av
,
Pk
and
Pk/Av
ratio:
1. Select "Trace & Statistics" in the "Mode" dialog box of the "Sensor" menu
2. Select "T'slot" mode in the "Display" control panel of the Trace display
3. Set the timing parameters in the "Config T'slot" control panel of the Trace display
The timeslot is indicated graphically in the trace display. The power values within
the selected gate are indicated in the info panel.
Fig. 4-37: Measurement within a timeslot (Trace&Statistics mode/Graph display mode)
To display the results numerically continue as follows:
1. Select "Dig" or "Ana&Dig" display in the "Window" dialog of of the "Windows" menu
2. Open the "Function&Unit" dialog box of the of the "Measurement "menu.
3. Select the timeslot, the main value ("Eval"), and "Aux. = Normal"
The main value together with the auxilliary values is indicated numerically. .
Fig. 4-38: Measurement within a timeslot (Trace&Statistics mode/ Dig display mode)
In "Dig"/"D&A" display modes, the results can be correlated to any other measured
value ("Function")
If you want to combine the measurement results of different channels by using func-
tions, not all combinations are supported. The "Ratio" function calls for measurands of
the same dimension, e.g. two power levels or two "Pk/Av" ratios. The impedance mea-
surement functions even require identical measurands, e.g. two average power values
but not one average power value and one peak power value. If both channels are set
to timeslot measurements, only identical timeslot numbers are supported.