Instrument Functions
NRP2
120
User Manual 1173.9140.02 ─ 07
Trace 1 / 2
Selects the trace (waveform) to be configured.
Up to two traces can be selected. This is also true if only one measurement channel
(sensor) is present. In this case, one trace for example can represent the average
envelope power and the other trace the peak or random power (available for the R&S
NRP-Z8x wideband sensors only).
Function
Selects the measurement function for the "Trace" mode. Only "Single" and "Ratio"
functions are available for trace measurements. Refer to
description.
Note:
It is essential that the sensors of both channels support trace measurement, if
the "Ratio" function is selected.
Prim. / Sec. Channel
Selects the primary and secondary channel.
Eval.
Select the type of display. Since each display point (pixel) represents a time interval
and thus usually many samples, the R&S
NRP-Z8x wideband power sensors allow you
to select what is to be displayed.
"Average"
Average power value, resulting in a flicker-free display and smooth
trace.
Note:
Average is fixed for all multipath power sensors.
"Random"
Power of a randomly selected sample, i.e. a realistic display with sig-
nal details.
"Peak"
Highest power measured (peak power).
Max. Hold
Displays the highest value measured for each point (pixel) of the selected display type
("Average", "Random ", "Peak").
"On / Off"
Starts/stops "Max Hold".
"Reset"
Resets the display and restarts recording.
4.5.4 Gates display mode
In the "Gate" mode, four separately configurable gates for signal evaluation are availa-
ble. The active gate is indicated in the window, the info panel shows the measurement
results and borders for this gate and the selected trace (waveform) . The settings for
"Trig"gering, "Filt"ering, "Function", and "Unit" are taken from the "Trace" display mode
and cannot be changed.
► Select "Gates" in the control panel opened via the Window
→
Graph
→
"Display" dia-
log box ("Trace&Statistics" sensor mode).
Displaying traces