Remote Control - Commands
NRP2
198
User Manual 1173.9140.02 ─ 07
Query:
*PSC?
The query form
*PSC?
reads out the contents of the power-on-status-clear flag. The
response can be 0 or 1.
Parameters:
<Function>
0 | 1
0
Causes the contents of the status registers to be preserved.
Thus a service request can be generated when switching on the
instrument, if the status registers ESE and SRE are suitably con-
figured.
1
Resets the registers.
*RCL
<value>
R
e
C
a
L
l calls the device state which has been stored with the
*SAV
command under
the specified number. The effect of
*RCL
0 to 19 is the same as if one of the setups 0
to 19 had been selected in manual control.
*RCL
0 resets the device to the default
state.
Parameters:
<Value>
1 to 19
Available recall numbers for device setups.
Usage:
Event
*RST
R
e
S
e
T
sets the instrument to a defined default status. The command essentially corre-
sponds to command
SYSTem:PRESet
and to pressing (PRE)SET.
The following table documents the difference.
SYSTem:PRESet
acts like the frontpanel
key (PRE)SET.
Table 6-1: Difference between commands *RST und SYSTem:PRESet
INITiate:CONTinuous
SENSe:AVERage:TCONtrol
*RST
OFF
REPeat
SYSTem:PRESet
ON
MOVing
The following table lists all preset values
Common Commands to IEEE 488.2