Instrument Functions
NRP2
101
User Manual 1173.9140.02 ─ 07
Prim. / Sec. Channel
Selects the primary channel for single channel measurement and selects the primary
and secondary channel for two channel measurements.
Fig. 4-31: Selection of the primary channel
Incident / Reflected
Selects the channel referenced to the incident power or the reflecting power. These
parameters appear instead of "Prim. / Sec. channel " if impedance measurement func-
tions such as "SWR", "RCoeff" and "RLoss" are selected.
4.4.3.2
Measuring power in gates
For time-gated power measurements the R&S
NRP2 power meter offers two modes:
●
T'gate Av
●
Trace & Statistics
T'gate Av
In this mode, the average power (
Av
) within a user-definable gate can be measured.
The time resolution is high and corresponds to the sampling rate of the sensor.
To configure the measurement:
1. Select "T'gate Av" in the "Mode" dialog box of the "Sensor" menu
2. Set the timing parameter for the gates
3. Select the gate number for the measurement in the "Function & Unit" dialog box of
the "Measurement" menu.
The average power within the selected gate is indicated.
Configuring measurements