Remote Control - Commands
NRP2
258
User Manual 1173.9140.02 ─ 07
OUTPut:RECorder<[1]..2>[:STATe]
<state>
Switches the analog outputs (OUT1 and OUT2/TRIG on rear panel)
ON
or
OFF
. The
magnitude of the output voltage (0 V to 3.3 V) depends on the measured power of the
assigned calculate block (
OUTP:REC:FEED
) and the configuration of the characteristic
(
OUTP:REC:LIM
). As in manual operation, this setting is coupled to other operating
modes of the outputs. Only one mode can be active at any one time:
Connector
Analog output
(
OUTP:REC
)
TTL output
(
OUTP:TTL
)
Off
Ext. Trigger In
OUT1
x
x
x
-
OUT2/TRIG
x
-
-
x
Parameters:
<state>
ON | OFF
*RST:
output 1: ON; output 2: OFF
OUTPut:ROSCillator[:STATe]
<state>
The command switches the optional RF generator (
ON
) or (
OFF
).
Parameters:
<state>
ON | OFF
*RST:
OFF
OUTPut:TTL:ACTive
<string>
Specifies whether the value set via
OUTP:TTL:LVOL (LOW)
or via
OUTP:TTL:HVOL
(HIGH)
should be applied if the measured value causes a limit violation at the TTL
output.
Parameters:
<string>
LOW | HIGH
*RST:
LOW
OUTPut:TTL:FAIL
<mode>
Specifies the value which should be applied if the measured value causes a limit viola-
tion at the TTL output.
Note:
The
OUTP:TTL:ACT
and
OUTP:TTL:FAIL
commands are identical, except for
the opposite meaning of the parameters.