Remote Control - Commands
NRP2
268
User Manual 1173.9140.02 ─ 07
[SENSe<[1]..4>]:CORRection:DCYCle:STATe
<state>
Switches duty cycle correction for the measured-value on or off.
Parameters:
<state>
ON | OFF
*RST:
depending on sensor
[SENSe<[1]..4>]:CORRection:FDOFfset[:INPut][:MAGNitude]?
Queries the frequency dependent offset. The frequency-dependent measured-value
correction is set by selecting a table with the
[SENSe<[1]..4>]:CORRection:FDOTable[:SELect]
and
[SENSe<[1]..4>]:CORRection:FDOTable:STATe ON
commands. The respec-
tive frequency-dependent correction factor is determined by means of the active tables
(if necessary by interpolation) and can be queried with
[SENSe<[1]..4>]:CORRection:FDOFfset[:INPut][:MAGNitude]?
(fre-
quency-dependent offset).
Unit:
DB
Usage:
Query only
[SENSe<[1]..4>]:CORRection:FDOTable[:SELect]
<table_name>
The name of the current frequency-dependent offset table can be set or queried by the
[SENSe<[1]..4>]:CORRection:FDOTable[:SELect][?]
command. The
R&S
NRP2 can manage up to ten tables (
MEMory
commands). The frequency-
dependent offset tables consist of two columns; the first containing the frequencies, the
second the associated correction values. When the R&S
NRP2 is informed about the
signal frequency with command
[SENSe<[1]..4>]:FREQuency[:CW|:FIXed]
<float_value>
, an appropriate correction factor is determined by means of the
table. If the table was activated with
[SENSe<[1]..4>]:CORRection:FDOTable:STATe[?] ON
, the measurement
result obtained by the sensor is multiplied by this factor.
Error messages:
-256
"File name not found"
: An invalid <string> parameter was transmitted.
Parameters:
<table_name>
string
*RST:
Settings are not changed in case of *RST.
[SENSe<[1]..4>]:CORRection:FDOTable:STATe
<state>
Activates and deactivates the current frequency-dependent offset table. The current
table is selected with the
[SENSe<[1]..4>]:CORRection:FDOTable[:SELect]
command. Only one table can be active at a time.