Remote Control - Commands
NRP2
259
User Manual 1173.9140.02 ─ 07
Parameters:
<mode>
HIGH
The value set via
OUTP:TTL:LVOL
is applied if the measured
value causes a limit violation at the TTL output.
LOW
The value set via
OUTP:TTL:HVOL
is applied if the measured
value causes a limit violation at the TTL output.
*RST:
HIGH
OUTPut:TTL:FEED
<calculate_block>
Sets the calculate block whose limit handling controls the TTL voltage. If the measured
value of this calculate block violates a limit and limit monitoring is active, the voltage
specified in
(first line) is output.
Parameters:
<calculate_block>
string
Range:
"CALC1:LIM","CALC2:LIM","CALC3:LIM","CALC4:L
IM"
*RST:
"CALC1:LIM"
OUTPut:TTL:HVOLtage
<high_voltage>
Sets the high voltage for the TTL output.
Parameters:
<high_voltage>
float_value
Range:
0 V to 3.3 V
*RST:
3.3 V
Default unit: V
OUTPut:TTL:LVOLtage
<low_voltage>
Sets the low voltage for the TTL output.
Note:
The value for the lower TTL level may be greater than the value for the upper
TTL level.
Parameters:
<low_voltage>
float_value
Range:
0 V to 3.3 V
*RST:
0 V
Default unit: V
OUTPut