Remote Control - Commands
NRP2
283
User Manual 1173.9140.02 ─ 07
Parameters:
<rlevel>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: DBM
[SENSe<[1]..4>]:STATistics:TIME
< time_interval>
Determines, together with the
SENS:STAT:OFFS
command, the time interval in which
the power for the statistical evaluation is measured.
SENS:STAT:TIME
determines the
length of the interval (see also
[SENSe<[1]..4>][:POWer]:TGATe<[1]..4>:
Parameters:
< time_interval>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
[SENSe<[1]..4>]:TIMing:EXCLude:STARt
<exclude_start>
Sets a time that is to be excluded at the beginning of the integration period.
Integration period
EXCLude:STARt
EXCLude:STOP
Power
Time
Fig. 6-15: Effect of commands SENS<[1]..4>:TIM:EXCL:STAR and :STOP
Parameters:
<exclude_start>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
[SENSe<[1]..4>]:TIMing:EXCLude:STOP
<exclude_stop>
Sets a time that is to be excluded at the end of the integration period. See
Effect of commands SENS<[1]..4>:TIM:EXCL:STAR and :STOP
.
SENSe (Sensor Commands)