Remote Control - Commands
NRP2
320
User Manual 1173.9140.02 ─ 07
Usage:
Event
SYSTem:BEEPer:NOTify:KEY
<state>
Enables (
ON
) or disables (
OFF
) acoustic key clicks, respectively.
Parameters:
<state>
ON | OFF
*RST:
OFF
SYSTem:BEEPer:STOP
Stops the output of an audible signal.
Usage:
Event
SYSTem:BEEPer:TIME
<nrf>
Sets the length in milliseconds of the audible signal output with
SYS:BEEP
.
Parameters:
<nrf>
int_value
Range:
1...60
*RST:
This setting is not changed by *RST.
Default unit: S
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
<nr1>
Sets the address with which the R&S
NRP2 can be addressed via the IEC/IEEE bus.
The address is factory-set to 20 and is not changed by a reset.
Parameters:
<nr1>
int_value
Range:
0...30
*RST:
This setting is not changed by *RST.
Usage:
Event
SYSTem:COMMunicate:INET:MODE
<mode>
Selects the mode for transmitting the special control sequences via ethernet according
to draft IEEE 1174.
Parameter
Mode
0
no control sequences (abbr.: 1174.0 mode)
1
the IEC/IEEE 488.1 mode (abbr.: 1174.1 mode)
2
the IEC/IEEE 488.2 mode (abbr.: 1174.2 mode)