Instrument Functions
NRP2
157
User Manual 1173.9140.02 ─ 07
0 V Equivalent - Analog ← Analog Out ← Mode
Entry of the measurement value that corresponds to the output voltage of 0
V.
Remote command:
OUTPut:RECorder<[1]..2>:LIMit:LOWer
OUTPut:RECorder<[1]..2>:LIMit:LOWer:POWer
OUTPut:RECorder<[1]..2>:LIMit:LOWer:RATio
3 V Equivalent - Analog ← Analog Out ← Mode
Entry of the measurement value that corresponds to the output voltage of 3
V.
Remote command:
OUTPut:RECorder<[1]..2>:LIMit:UPPer
OUTPut:RECorder<[1]..2>:LIMit:UPPer:POWer
OUTPut:RECorder<[1]..2>:LIMit:UPPer:RATio
Trigger In ← Mode
The connector is used as input. An external trigger signal can be applied.
Fig. 4-88: I/O 2 mode / Trigger In
Use the external trigger signal, e.g. to start the measurement in all sensors simultane-
ously. Provided the R&S
NRP2 is configured accordingly, it passes the signal directly
through to the sensors. Refer to
chapter 4.2.6, "Trigger settings"
mation on the respective settings.
Remote command:
The state must be set to
OFF
.
OUTPut:RECorder<[1]..2>[:STATe]
4.7.2.3
Voltage at the analog outputs
The voltage at the analog outputs does not track the power fed to the sensor continu-
ously but rather at discrete points of time. Every time the sensor has transmitted a new
measured value to the base unit, i.e. after an aquisition has been completed (
ter 4.2.3, "Filter / averaging"
on page 79), the analog output voltage is adjusted to the
new value.
In the ContAv mode, you can influence the rate at which the voltage is updated by set-
ting the aperture time to a convenient value, see
chapter 4.2.1.2, "Continuous average-
on page 66. The time between two updates of the output volt-
age is approximately twice the length of the aperture time. The lower limit of this length