[SENSe<[1]..4>][:POWer][:AVG]:RANGe
<range>
Selects a measurement range in which the corresponding sensor is to perform a mea-
surement. This setting will only become effective if
SENS<[1]..4>:RANG:AUTO OFF
is set.
Parameters:
<range>
0 | 1 | 2
*RST:
depending on sensor
[SENSe<[1]..4>][:POWer][:AVG]:RANGe:AUTO
<state>
Activates/deactivates the automatic selection of a measurement range.
Parameters:
<state>
ON | OFF
*RST:
depending on sensor
[SENSe<[1]..4>][:POWer][:AVG]:RANGe:CLEVel
<level>
Sets the cross-over level. Shifts the transition ranges between the measurement
ranges. This may improve the measurement accuracy for special signals, i.e. signals
with a high crest factor.
Unit:
DB | PCT
Parameters:
<clevel>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: DB
[SENSe<[1]..4>][:POWer][:AVG]:SMOothing:STATe
<state>
Activates/deactivates digital lowpass filtering of the sampled video signal.
The problem of instable display values due to a modulation of a test signal described
under
SENS<[1]..4>:SAMP
can also be eliminated by lowpass filtering of the video
signal. The lowpass filter eliminates the variations of the display even in case of unperi-
odic modulation and does not require any other setting.
If modulation is periodic, setting the aperture time is the better method, since it allows
shorter measurement times.
Parameters:
<state>
ON | OFF
*RST:
depending on sensor