
Instrument Functions
NRP2
78
User Manual 1173.9140.02 ─ 07
Table (state)
Activates/deactivates the frequency-dependent offset correction specified in the
selected table.
Remote command:
[SENSe<[1]..4>]:CORRection:FDOTable:STATe
Edit Table ...
Opens a dialog box for editing the selected offset correction table, see
"Table editor for offset tables"
Remote command:
S Parameter Device
Activates the S-parameter data set stored in the sensor for a two-port that is connected
to the sensor input (output).
Remote command:
[SENSe<[1]..4>]:CORRection:SPDevice:STATe
4.2.2.2
Table editor for offset tables
The R&S
NRP2 provides a list editor for defining the frequency / offset value pairs.
Two table-editor modes are available:
●
Selection mode
In this mode, using the cursor keys
⇧
⇩
⇦
⇨
, you can select the number field you
want to "Edit", "Insert" or "Delete".
Note:
A long list can not be displayed in full length in the dialog box. The entries
are, therefore, viewed in a “window” which automatically scrolls up or down, if the
cursor moves outside the visible section. The position of the currently active field is
shown by the scroll bar.
●
Edit mode
When you press the "Edit" rocker switch the value in the selected field can be edi-
ted. In this mode, the cursor keys cannot be used for selection, but can be used
“normally” to edit values in the entry field.
► To open the table editor, select "Sensor" > "Mode" > "Offset", and then "Edit
Table..." in the "Offset" dialog box.
Data acquisition and parameters
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com