Remote Control - Commands
NRP2
314
User Manual 1173.9140.02 ─ 07
Bit No.
Meaning
8
Sensor D please zero
The zero correction for sensor D is no longer correct and should be repeated.
9 to 14
Not used
15
Bit 15 will never be used.
The sensor power bit is set when error -230, "Data corrupt or stale", or -231, "Data
questionable", occurs in the respective sensor.
6.13.18 Questionable Window Status Register
The CONDition register contains information as to whether the displayed data or the
power calculated by the calculate blocks is questionable.
The Questionable Window Status Register can be read by the following commands:
STATus:QUEStionable:WINDow[:SUMMary]:CONDition?
STATus:QUEStionable:WINDow[:SUMMary][:EVENt]?
Table 6-30: Meaning of bits used in the Questionable Window Status Register
Bit No.
Meaning
0
Not used
1
Window 1 Power
The measured values returned by Calculate block 1 are corrupt.
2
Window 2 Power
The measured values returned by Calculate block 2 are corrupt.
3
Window 3 Power
The measured values returned by Calculate block 3 are corrupt.
4
Window 4 Power
The measured values returned by Calculate block 4 are corrupt.
5
Window 5 Power
The measured values returned by Calculate block 5 are corrupt.
6
Window 6 Power
The measured values returned by Calculate block 6 are corrupt.
7
Window 7 Powe
r
The measured values returned by Calculate block 7 are corrupt.
8
Window 8 Power
The measured values returned by Calculate block 8 are corrupt.
9 to 14
Not used
15
Bit 15 will never be used.
STATus
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com