Remote Control - Commands
NRP2
229
User Manual 1173.9140.02 ─ 07
Table 6-10: Value ranges, depending on the unit
Unit of mea-
sured value
W
DBM
DBUV
DPCT
DB
O
Value range
-9.99999E5
..
9.99999E5
-120
..
90
-13.0103
..
196.897
-100
..
999.99
-200
..
200
-1E20
..
1E20
CALCulate<[1]..8>:LIMit:LOWer[:DATA]:POWer
<lower_limit>
Sets a lower limit for the measured power. The R&S
NRP2 can respond when the val-
ues drop below this limit (see other
CALCulate<[1]..8>:LIMit
commands).
Parameters:
<lower_limit>
float_value; units: W | DBM | DBUV
Range:
-120 DBM to 90 DBM
*RST:
0 DBM
Default unit: DBM
CALCulate<[1]..8>:LIMit:LOWer[:DATA]:RATio
<ratio>
Sets a lower limit for the measured power ratios. The R&S
NRP2 can respond when
the ratios drop below this limit (see other
CALCulate<[1]..8>:LIMit
commands).
Parameters:
<ratio>
float_value; units: DB | DBCT | O
Range:
-200 DB to 200 DB
*RST:
0 DB
Default unit: DB
CALCulate<[1]..8>:LIMit:LOWer:STATe
<state>
Switches the monitoring function for theupper and lower limit on or off.
Parameters:
<state>
ON | OFF
*RST:
OFF
Usage:
Setting only
CALCulate<[1]..8>:LIMit:STATe
<state>
Switches the monitoring function for the upper limit on or off.
Parameters:
<state>
ON | OFF
*RST:
OFF
Usage:
Setting only