Instrument Functions
NRP2
134
User Manual 1173.9140.02 ─ 07
4.5.7.3
T'Slot diagram settings
► Configure timeslot measurement in the "Timeslot Config" dialog box which is
opened via the Window
→
Graph
→
"Display"
→
"T'slot" control panel ("Trace&Statis-
tics" sensor mode).
Fig. 4-60: Timeslot Config dialog box
No of Timeslots
Defines the number of subsequent timeslots within a frame, i.e. a period of the signal.
Remote command:
[SENSe<[1]..4>][:POWer]:TSLot:COUNt
Nominal Width
Defines the length of a timeslot.
Example:
If measurements are performed on TDMA signals (GSM/EDGE, PDC, NADC, PHS
etc), the nominal timeslot length must be entered here, i.e. the frame length divided by
the number of timeslots.
Table 4-10: GSM signal
Frame length:
4.615 ms
No. timeslots:
8
⇒
Nominal width:
4.615 / 8 ms = 576.875
μ
s
Remote command:
[SENSe<[1]..4>][:POWer]:TSLot:[AVG]:WIDTh
Excluded from Start / End
Defines the intervals at the start or end of the timeslot which are excluded from the
measurement. Measurement can thus be restricted to the relevant part of a timeslot.
Remote command:
[SENSe<[1]..4>]:TIMing:EXCLude:STARt
[SENSe<[1]..4>]:TIMing:EXCLude:STOP
Fence
Activates the "Fence" function.
Displaying traces