Limit warnings may be configured separately for each window. Therefore, the check-
box "Limit Warning" may show a grayed checkmark, if the settings differ for different
windows.
Beep - Overload
Activates an acoustic feedback when the sensor is overloaded (overload condition, see
chapter 4.8, "Messages and alarms"
Remote command:
Beep - Key Click
Activates an acoustic alarm to indicate when a key is pressed.
Remote command:
Beep - Limit Warning
Activates an acoustic alarm when there is a violation of user-defined limits.
Activation applies to all windows. If a warning beep in conjunction with a violation of
user-defined limits is not desired, it can explicitly and exclusively for any window be
switched off in the limits dialog, see
chapter 4.4.5, "Limit Monitoring"
Display Update
Selects the display refresh rate.
At very low ambient temperatures, the display becomes sluggish. To improve readabil-
ity under these circumstances, you can reduce the display update rate.
"Normal"
standard display update rate.
"Slow"
preferred setting rate at very low ambient temperatures.
"Frozen"
for interruption-free update of analog outputs, see
"Voltage at the analog outputs"
on page 157. The meter indicates
"frozen" in the info line of the results window.
Remote command:
Brightness
Set the display brightness.
Remote command:
4.7.6 Test measurements
Since the scope of testing is limited, test measurements cannot replace a regular reca-
libration of the sensor.
System settings