Annex
NRP2
402
User Manual 1173.9140.02 ─ 07
E4418B/E4419B command
Compatibility information
SENSe
:AVERage:COUNt:AUTO:
SDETect
:DETector:FUNCtion
:MRATe
:SPEed
:V2P
Is not implemented in the R&S
NRP2.
SENSe[1|2]:CORRection
:CSET[1]|CSET2[:SELect]
and
SENSe[1|2]:CORRection
:CSET[1]|CSET2:STATe
The R&S
NRP2 does not recognize any tables with frequency-
dependent calibration values since the sensors have to be calibrated
only once in production. For this reason, there is no compliance with
CSET1
. Tables with frequency-dependent offset correction values
can be defined and addressed via
:CSET2
(Agilent and R&S
NRP2)
or
:FDOTable
(R&S
NRP2 only).
SENSe[1|2]:CORRection:
GAIN[1]?
Has the same effect as
SENSe[1..2]:CORRection:CFACtor?
SENSe[1|2]:CORRection:
GAIN2[?]
Has the same effect as
SENSe[1..2]:CORRection:OFFSet[?]
.
With
:GAIN2
, no unit may be specified. Unit 1 is valid.
SENSe[1|2]:CORRection:
LOSS2[?]
Corresponds to the reciprocal of
SENSe[1..2]:CORRection:OFFSet[?]
. With
:LOSS2
, no unit
may be specified. Unit 1 is valid.
SENSe[1|2]:CORRection:
GAIN3[?]
Has the same effect as
SENSe[1..2]:CORRection:DCYCle[?]
.
With
:GAIN3
, no unit may be specified. Unit 1 is valid.
SENSe[1|2]:CORRection:
GAIN4?
Has the same effect as
SENSe[1..2]:CORRection:FDOFfset[:INPut][:
MAGNitude]?
SENSe:SWEep:OFFSet:TIME[?]
Is implemented in the R&S
NRP2 and has the same effect as
SENSe:TGATe:OFFSet:TIME[?]
SENSe:SWEep:TIME[?]
Is implemented in the R&S
NRP2 and has the same effect as
SENSe:TGATe:TIME[?]
SYSTem
:COMMunicate:SERial
:HELP:HEADers?
:LOCal
:REMote
:RINTerface
:RWLock
Is not implemented in the R&S
NRP2.
TRACe:STATe
Is not implemented in the R&S
NRP2.
TRACe:DATA?
Is not implemented in the R&S
NRP2 in this form. Data of the trace
mode can only be queried via the high-level commands (
FETCh?
,
READ?
,
MEAS?
) and
CALC:DATA?
.
TRACe:UNIT?
The
UNIT
node is only globally available in the R&S
NRP2 and
affects all measurement results (including trace data) and the unit-
less command parameters of the commands
CALC:LIM:LOW
,
CALC:LIM:UPP
,
DISP:MET|:ANAL:LOW
,
DISP:MET|:ANAL:UPP
,
OUTP:REC:LIM:LOW
and
OUTP:REC:LIM:UPP
.