Annex
NRP2
380
User Manual 1173.9140.02 ─ 07
The second parameter may assume the values 1, 2, 3 and 4. It specifies the number of
places up to which the measurement result should be noise-free. With linear units, the
setting corresponds to the number of significant places that are noise-free; with loga-
rithmic units, 1, 2, 3 and 4 correspond to a resolution of 1, 0.1, 0.01 and 0.001 (number
of decimal places). With a
<resolution>
of
3, two decimal places are thus noise-free
with logarithmic display. The default setting is
3.
The third parameter selects a sensor. It has the syntax (@n), where n can assume the
values 1 to 4. The default setting is (@1).
A.4.1.2
Relative measurements
If the measurement results are to be output not as absolute values, but relative to a
reference value, the keyword
:RELative
can be added to the measurement com-
mand. The result is then divided by a value that was previously set by means of
CALC:REL:POW
or
CALC:REL:AUTO ONCE
.
MEAS?
CALC:REL:AUTO ONCE
MEAS:REL?
The first
MEAS
command determines the reference value, which is acquired and stored
with
CALC:REL:AUTO ONCE
. The second measurement command
MEAS:REL?
and
all other
MEAS:REL?
commands will divide the absolute measured value by the stored
reference value.
CALC:REL:POW <float_value>
is used to define such a reference value.
CALC:REL:POW 0 DBM
MEAS:REL?
All measured values are then divided by 0 dBm (1 mW) prior to being transferred to the
user.
The following section describes an extension of the
MEAS
command in which the result
no longer carries a power unit, but represents a power ratio. The reference value also
must have a power unit and be set with
CALC:REL:RAT
. The R&S
NRP2 stores the
reference values for power values and power ratios independently of each other and
automatically uses the correct reference value depending on the unit of the measure-
ment result.
The
CALC:REL
command is available for entering the reference value. It assumes a
numeric parameter whose unit is determined from the current device settings. This
command is provided only for reasons of compatibility and should not be used unless
avoidable.
A.4.1.3
Processing the results of two sensors
The R&S
NRP2 offers measurement commands for processing the results of two sen-
sors. The calculated result can also be partly referenced to a stored reference value.
Programming Measurements