CMU-K20/K21/K22/K23/K24
Non Signalling: Power Measurements
1115.6088.12 4.21
E-11
the
Connection Control
menu
(see section
Table-Oriented
Version
on p.
4.73
ff.)
Attenuation
Setting for the attenuation of the input level
(Normal, Low
Noise, Low Distortion)
as set in the
Analyzer
tab of the
Connection Control
menu
(see section
Table-Oriented
Version
on p.
4.73
ff.)
Freq. Offset
Frequency offset compared to the nominal channel frequency
Chan./Trig. Slot Offs.
RF channel and trigger slot offset (see
Trig. Slot Offset
hotkey on p.
4.18
)
2
nd
parameter
line
The second parameter line contains the following marker values:
Level and time of reference marker
Level and time of delta marker 1 (setting
absolute
) or difference
from reference marker (setting
relative
)
Level and time of delta marker 2 (setting
absolute
) or difference from reference
marker (setting
relative
)
Info box
The info box contains the following settings:
Sym. Offset
Time Delay
set by means of the
Display/Marker
softkey: Number
of symbols that the burst is shifted with respect to the time axis
and the tolerance template. In
Signalling
mode, the measured
timing advance error is displayed instead.
Statistic Count
Number of bursts per statistics cycle, as set in the
Control
tab of
the
Power Configuration
menu.
In addition, the following scalar results are indicated:
Avg Burst Power
Average burst power, depending on the display mode set (see
upper right corner of the diagram).
TSC detected
Training sequence of the measured RF burst
(GSM – 0 to 7 |
Dummy | ---
.
Out of Tolerance
Relative number of bursts that are out of the tolerances defined
by the limit lines.
Burst Matching
Error message if the displayed curve is out of tolerance.
Remote control
Settings are retrieved using the query corresponding to the setting command
(setting command with appended question mark).
For scalar measurement results:
READ[:SCALar]:POWer[:NORMal][:GMSK]?
CALCulate[:SCALar]:POWer[:NORMal][:GMSK]
:LIMit:MATChing?
FETCh[:SCALar]:POWer[:NORMal][:GMSK]?
SAMPle[:SCALar]:POWer[:NORMal][:GMSK]?
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...