![Rohde & Schwarz CMU-K20 Operating Manual Download Page 409](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804409.webp)
CMU-K20/21/22/23/24 Common: SPECtrum:MODulation
1115.6088.12 6.105
E-12
CONFigure:SPECtrum:MODulation[:GMSK]:CONTrol:REPetition
Test cycles
CONFigure:SPECtrum:MODulation:EPSK:CONTrol:REPetition
<
Repetition
>
,<StopCondition>
,
<Stepmode>
<Repetition>
Description of parameters
Def. value
Def. unit
CONTinuous |
SINGleshot |
1 to 10000
Continuous measurement (until
STOP
or
ABORT)
Single shot measurement (until
Status = RDY
)
Multiple measurement
(
counting
, until
Status = STEP | RDY)
SING –
<StopCondition>
Description of parameters
Def. value
Def. unit
SONerror |
NONE
Stop measurement in case of error
(stop on error)
Continue measurement even in case of error
NONE –
<Stepmode>
Description of parameters
Def. value
Def. unit
FW vers.
STEP |
NONE
Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode
NONE –
V1.20
Description of command
This command defines the number of test cycles, the stepping mode and, if required, a stop condition for the
measurement.
Note:
In the case of READ commands (
READ:
...), the <Repetition> parameter has no effect; the measure-
ment is always stopped after a single shot.
DISPlay:SPECtrum:MODulation[:GMSK]:CONTrol:GRID
<Enable>
DISPlay:SPECtrum:MODulation:EPSK:CONTrol:GRID
<Enable>
Grid
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
Switch on grid lines
Switch off grid lines
ON – V1.20
Description of command
This command switches the grid lines in the test diagrams on or off.
CONFigure:SPECtrum:MODulation[:GMSK]:CONTrol:MPOint<nr>:ENABle <Enable>
CONFigure:SPECtrum:MODulation:EPSK:CONTrol:MPOint<nr>:ENABle <Enable>
Enable/Disable Measurement Points
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
Switch on measurement point <nr>
Switch off measurement point <nr>
ON – V1.20
Description of command
This command switches the measurement at the individual frequency points no. 1 to 11 (numbered by the nu-
meric suffix <nr>) on or off. Each number denotes a pair of frequency points symmetric to the carrier, <nr>=1
corresponding to
±
0.1 MHz, <nr>=11 to
±
1.8 MHz.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...