Common: Symbolic Status Event Register Evaluation
CMU-K20/21/22/23/24
1115.6088.12 6.34
E-12
CONFigure:IQIF:RXPath
<Path>
RX Path
<Path>
Description of parameters
Def. value
Def. unit
FW vers.
BYP
BYIQ
XOIO
IOIO
IOXO
Bypass
Bypass w. I/Q-IF OUT
I/Q IN/OUT
IF IN_I/Q IN/OUT
IF IN/OUT
BYP –
V3.10
Description of command
This command selects the RX signal path, leaving the TX path (see command
CONFigure:IQIF:TXPath
below) unchanged but adapting the I/Q-IF test scenario to the new RX/TX path combination: If the combination
corresponds to a predefined scenario, then
CONFigure:IQIF:RXTXcombined
is set to the predefined
scenario; otherwise it is set to
UDEF
.
CONFigure:IQIF:TXPath
<Path>
TX Path
<Path>
Description of parameters
Def. value
Def. unit
FW vers.
BYP
BYIQ
XOIO
IOIO
IOXO
Bypass
Bypass w. I/Q-IF OUT
I/Q IN/OUT
IF IN_I/Q IN/OUT
IF IN/OUT
BYP –
V3.10
Description of command
This command selects the TX signal path, leaving the RX path (see command
CONFigure:IQIF:RXPath
above) unchanged but adapting the I/Q-IF test scenario to the new RX/TX path combination: If the combination
corresponds to a predefined scenario, then
CONFigure:IQIF:RXTXcombined
is set to the predefined
scenario; otherwise it is set to
UDEF
.
IQIF:DEFault <Enable>
Default Settings
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
The parameters are set to their default values
Some or all parameters differ from the default values
ON
– V3.10
Description of command
If used as a setting command with the parameter
ON
this command sets all parameters of the subsystem to their
default values (the setting
OFF
causes an error message).
If used as a query the command returns whether all parameters are set to their default values
(ON)
or not
(OFF)
.
Symbolic Status Event Register Evaluation
The following commands are used to retrieve the events reported in function groups
GSM400/850/900/1800/1900-MS Non Signalling
and
Signalling
; see section
Symbolic Status Event
Register Evaluation
in Chapter 5 of the CMU operating manual.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...