Common: SPECtrum:SWITching CMU-K20/21/22/23/24
1115.6088.12 6.114
E-12
CONFigure:SPECtrum:SWITching[:GMSK]:CONTrol:MPOint<nr>:ENABle <Enable>
CONFigure:SPECtrum:SWITching:EPSK:CONTrol:MPOint<nr>:ENABle <Enable>
Enable/Disable Measurement Points
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
Switch on measurement point <nr>
Switch off measurement point <nr>
ON – V1.20
Description of command
This command switches the measurement at the individual frequency points no. 1 to 4 (numbered by the numeric
suffix <nr>) on or off.
CONFigure:SPECtrum:SWITching[:GMSK]:CONTrol:AMPoints:FREQuency <F1>, <F2>
CONFigure:SPECtrum:SWITching:EPSK:CONTrol:AMPoints:FREQuency <F1>, <F2>
Additional Measurement Points
<F1>, <F2>
Description of parameters
Def. value
Def. unit
FW vers.
0 MHz to 1.8 MHz |
OFF |
ON
Frequency offset of additional meas. points
Measurement at additional point switched off
Meas. switched on again at previously defined
offset frequencies (or 0 MHz, if no frequ. defined)
OFF MHz V3.40
Description of command
This command defines up to 2 pairs of additional measurement points for the spectrum due to switching meas-
urement. The measurement is performed at ± the specified offset frequencies. The results can be retrieved by
means of
FETCh:ARRay:SPECtrum:SWITching[:GMSK]:AMPoints:POWer?
(see p.
6.119
).
CONFigure:SPECtrum:SWITching[:GMSK]:CONTrol:DEFault
<Enable>
CONFigure:SPECtrum:SWITching:EPSK:CONTrol:DEFault
<Enable>
Default Settings
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
All the parameters of the subsystem correspond to their de-
fault values
At least one parameter of the subsystem differs from its de-
fault value
ON
– V2.00
Description of command
If used as a setting command with the parameter
ON
, this command sets all the parameters of the subsystem to
their default values (
OFF
causes an error message).
In the query format, the command returns
ON
if all the parameters of the subsystem correspond to their default
values, otherwise it returns
OFF
.
Test Configuration
The commands of the following subsystems configure the spectrum due to switching. They correspond
to the
due to Switching
sections in the
Spectrum Configuration
menu.
Subsystem SPECTrum:SWITching:LIMit:LINE
The subsystem
SPECtrum:SWITching:LIMit:LINE
defines the limit lines, i.e. the tolerance values for the
spectrum due to switching measurement.
The subsystem corresponds to the
due to Switching
sections
in the tab
Limit Lines
in the popup menu
Spectrum Configuration.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...