![Rohde & Schwarz CMU-K20 Operating Manual Download Page 464](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804464.webp)
Signalling: Receiver Quality
CMU-K20/21/22/23/24
1115.6088.12 6.160
E-12
CONFigure:RXQuality:BER<nr>:LIMit:DEFault <Enable>
Default Settings
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
All parameters are set to their default values
Some or all parameters differ from the default values
ON
– V2.00
Description of command
As a
setting command
with the setting
ON
this command sets all parameters of the subsystem to default values
(the setting
OFF
causes an error message).
As a query, this command reads out whether all parameters are set to default values
(ON)
or not
(OFF)
.
Measured Values – Subsystem RXQuality:BER
The subsystem
RXQuality:BER
measures and outputs the bit error rate and compares it with the
tolerance values. The subsystem corresponds to the measurement menus
Receiver Quality
for the
single shot type of measurement and measured values
(RFER, BER, BurstByBurst).
Scalar
Results
READ[:SCALar]:RXQuality:BER?
Start single shot measurement and return results
FETCh[:SCALar]:RXQuality:BER?
Read out meas. results (unsynchronized)
SAMPle[:SCALar]:RXQuality:BER?
Read out measurement results (synchronized)
Returned values for RFER
Value range
Def. value
Def. unit
ProgressTime,
ClassIIBits,
ClassIbBits,
FER,
CRCErrors,
Status
Statistical Result*
)
0.0% to 100.0%
0.000% to 100.000%
0.000% to 100.000%
0.000% to 100.000%
0 to 50000
INV | PASS | FAIL | TLOW | IMP
RUN | EFAI | EPAS | FAIL | PASS | THIG |
TLOW
NAN
NAN
NAN
NAN
NAN
INV
INV
%
%
%
%
–
–
–
Returned values for BER
Value range
Def. value
Def. unit
ProgressTime,
ClassIIBits,
ClassIbBits,
CRCErrors,
Status
Statistical Result*
)
0.0% to 100.0%
0.000% to 100.000%
0.000% to 100.000%
0 to 50000
INV | PASS | FAIL | TLOW | IMP
RUN | EFAI | EPAS | FAIL | PASS | THIG |
TLOW
NAN
NAN
NAN
NAN
INV
INV
%
%
%
–
–
–
Returned values for BBB
Value range
Def. value
Def. unit
ProgressTime,
BER,
CRCErrors,
Status
Statistical Result*
)
0.0% to 100.0%
0.000% to 100.000%
0 to 50000
INV | PASS | FAIL | TLOW | IMP
RUN | EFAI | EPAS | FAIL | PASS | THIG |
TLOW
NAN
NAN
NAN
INV
INV
%
%
–
–
–
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...