CMU-K20/21/22/23/24 Common:
POWer:FRAMe
1115.6088.12 6.53
E-12
Single
Result
READ[:SCALar]:POWer:FRAMe:FPOWer<nr>?
Start single shot measurement and return results
FETCh[:SCALar]:POWer:FRAMe:FPOWer<nr>?
Read out measurement results (unsynchronized)
SAMPle[:SCALar]:POWer:FRAMe:FPOWer<nr>?
Read out measurement results (synchronized)
Returned Values
Description of parameters
Def. value
Def. unit
FW vers.
–100 dBm to 100 dBm
Avg. power in slot <nr>
NAN
dBm
V2.15
Description of command
These commands are always queries. They start a measurement and return the average power in a particular
frame (numbered by <nr>=1 to 256, corresponding to frame 0 to frame 255 of the graphical user interface).
P/Slot Graph Results
READ:ARRay:POWer:FRAMe?
Start single shot measurement and return results
FETCh:ARRay:POWer:FRAMe?
Read out measurement results (unsynchronized)
SAMPle:ARRay:POWer:FRAMe?
Read out measurement results (synchronized)
Returned Values
Description of parameters
Def. value
Def. unit
FW vers.
–100 dBm to 100 dBm,
...
–100 dBm to 100 dBm
Avg. power in slot 0,
...,
Avg. power in slot Slot 255
NAN,
...,
NAN
dBm,
...,
dBm
V2.15
Description of command
These commands are always queries. They start a measurement and return all measurement results. The
returned list contains the average burst power of the mobile phone in 256 consecutive timeslots.
Subarray
Results
READ:SUBarrays:POWer:FRAMe?
Start single shot measurement and return results
FETCh:SUBarrays:POWer:FRAMe?
Read out measurement results (unsynchronized)
SAMPle:SUBarrays:POWer:FRAMe?
Read out measurement results (synchronized)
Ret. values per subrange
Description of parameters
Def. value
Def. unit
FW vers.
–100 dBm to 100 dBm,
...
–100 dBm to 100 dBm
Avg. power in first slot,
...,
Avg. power in last slot
NAN
...
NAN
dB
...
dB
V2.15
Description of command
These commands are always queries. They output the burst power versus time in a fixed ¼- bit pattern and in
the subranges defined by means of the
CONFigure:SUBarrays:POWer[:NORMal][:GMSK]:MPR
command.
In the default setting of the configuration command the
READ:SUBarrays..., FETCh:SUBarrays..., and
SAMPle:SUBarrays...
command group is equivalent to the
READ:ARRay..., FETCh:ARRay..., and
SAMPle:ARRay...
command group described above.
The
CONFigure:SUBarrays:POWer[:NORMal][:GMSK]:MPR
command defines a maximum of 32
subranges. If one of the statistical modes
(ARIThmetical, MINimum, MAXimum)
is set, only one value is
returned per subrange.
The calculation of
current, average, minimum,
and
maximum
results is explained in chapter 3 (cf.
display mode
).
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...