Common: SPECtrum:MODulation
CMU-K20/21/22/23/24
1115.6088.12 6.108
E-12
CONFigure:SPECtrum:MODulation[:GMSK]:LIMit:LINE:MODE[:UPPer]
<Mode>
CONFigure:SPECtrum:MODulation:EPSK:LIMit:LINE:MODE[:UPPer]
<Mode>
Limits on/off
<Mode>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
Switch on limit lines
Switch off limit lines
ON – V1.20
Description of command
This command switches all limit lines for the spectrum due to modulation measurement on or off.
CONFigure:SPECtrum:MODulation[:GMSK]:LIMit:LINE:DEFault
<Enable>
CONFigure:SPECtrum:MODulation:EPSK:LIMit:LINE:DEFault
<Enable>
Default Settings
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
All the parameters of the subsystem are set to default val-
ues
At least one parameter of the subsystem differs from its de-
fault value
ON
– V2.00
Description of command
If used as a setting command with
ON
, this command sets all the parameters of the subsystem to their default
values (
OFF
causes an error message).
In the query format, the command returns
ON
if all the parameters of the subsystem correspond to their default
values, otherwise it returns
OFF
.
Subsystem SUBarrays:SPECtrum:MODulation
The subsystem
SUBarrays:SPECtrum:MODulation
defines the measurement range and the type of
output values.
CONFigure:SUBarrays:SPECtrum:MODulation[:GMSK]
CONFigure:SUBarrays:SPECtrum:MODulation:EPSK
<Mode>,<Start>,<Samples>{,<Start>,<Samples>}
Definition of Subarrays
<Mode>
Description of parameters
Def. value
Def. unit
ALL |
ARIThmetical |
MINimum |
MAXimum |
IVAL,
Return all measurement values
Return arithm. mean value in every range
Return minimum value in every range
Return maximum value in every range
Return single interpolated value at <Start>
ALL –
<Start>
Description of parameters
Def. value
Def. unit
–1.8 MHz to 1.8 MHz,
Frequency of first point in current range
–1.8
MHz
<Samples>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 23
Number of samples in current range
23
–
V2.00
Description of command
This command configures the
READ:SUBarrays.., FETCh:SUBarrays..,
and
SAM-
Ple:SUBarrays:SPECtrum:MODulation
commands. It restricts the measurement to up to 32 subranges
where either all measurement results (the number of which is given by the second numerical parameter) or a
single statistical value is returned. The subranges are defined by the start time and the number of test points
which are located at fixed frequencies (see command
CONFig-
ure:SPECtrum:MODulation:LIMit:LINE<nr>
). If <Start> does not coincide with a test point then the range
will start at the next test point that is larger than <Start>.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...