CMU-K20/21/22/23/24 Signalling: Receiver Quality
1115.6088.12 6.171
E-12
Subsystem RXQuality:BLER:CONTrol
The subsystem
RXQuality:BLER:CONTrol
defines parameters controlling the scope of the BLER
measurement. The subsystem corresponds to the
Control
tab in the popup menu
Receiver Quality
Configuration.
CONFigure:RXQuality:BLER:CONTrol:REPetition
<Repetition>
,
<Stepmode>
Stop
Condition
<Repetition>
Description of parameters
Def. value
Def. unit
CONTinuous |
SINGleshot |
1 to 10000
Continuous measurement (until
STOP
or
ABORT)
Single shot measurement (until
Status = RDY
)
Multiple measurement
(
counting
, until
Status = STEP | RDY)
SING –
<StopCond>
Description of parameters
Def. value
Def. unit
NONE
Continue measurement even in case of error
NONE
–
<Stepmode>
Description of parameters
Def. value
Def. unit
FW vers.
STEP |
NONE
Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode
NONE –
V2.00
Description of command
This command determines the number of statistics cycles, the stop condition and the stepping mode for the
measurement. The second parameter is reserved for the limit check and has no effect at present.
Note:
In the case of READ commands (
READ:
...), the
<Repetition>
parameter has no effect; the
measurement is always stopped after a single shot
CONFigure:RXQuality:BLER:CONTrol:RLBCount
<Blocks>
RLC Block Count
<Blocks>
Description of parameters
Def. value
Def. unit
FW vers.
100 to 2000
Number of RLC blocks
2000
(blocks)
V3.10
Description of command
This command sets the number of RLC blocks to be sent and evaluated in a single shot BLER measurement.
Reference Level, Multislot
CONFigure:RXQuality:BLER:CONTrol:PDATa[:TCH]:MSLot:RLEVel
<Level>
<Level>
Description of parameters
Def. value
Def. unit
FW vers.
–
137 dBm to –27 dBm
–137 dBm to –10 dBm
–90 dBm to +13 dBm
RF1 reference level
RF2 reference level
RF3 OUT reference level
–85
–85
–85
dBm
dBm
dBm
V3.10
Description of command
This command defines the reference value for the individual downlink (BS) TCH signal levels used for the
multislot BLER test on packet data channels. See command
CONFigure:RXQuality:BLER:CONTrol:PDATa[:TCH]:MSLot:LEVel:INDividual
.
Slot Configuration: Individual (Multislot)
CONFigure:RXQuality:BLER:CONTrol:PDATa[:TCH]:MSLot:LEVel:INDividual
<Level_0>, ..., <Level_7>
<Level_n>
Description of parameters
Def. value
Def. unit
FW vers.
–127 dB to +127 dB
Power of CMU in timeslot no. n
0
dB
V3.10
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...