CMU-K20/-K21/-K22/-K23/-K24
General Settings
1115.6088.12
3.7
E-10
mum
and
Max./Min.
on the other hand, if the measurement extends over more
than one statistic count (repetition mode
Continuous,
measurement time longer
than one statistic count).
After evaluation of the different traces, the burst power is logarithmized and
plotted in a semi-logarithmic diagram.
The
Display Mode
is set in the
Control
tab of the configuration popup-menus
assigned to the measurement groups
Power, Modulation,
and
Spectrum.
Calculation of
average quantities
The
Average
traces in the
Power, Modulation,
and
Spectrum
menus are ob-
tained as follows:
Let c be the number of bursts forming one statistics cycle (one
Statistic Count)
and assume that n bursts have been measured since the start of the measure-
ment. In calculating the
Average
trace, the following two situations are distin-
guished:
n
≤≤≤≤
c
Single shot measurement or continuous measurement during the
first statistics cycle: At each test point,
Average
trace no. n is cal-
culated from
Average
trace no. n – 1 and
Current
trace no. n ac-
cording to the following recurrence:
)
,
,
1
(
)
(
1
)
1
(
1
)
(
c
n
n
Curr
n
n
Avg
n
n
n
Avg
K
=
+
−
−
=
The
Average
trace represents the arithmetic mean value over all
n bursts measured.
n > c
Continuous measurement after the first statistics cycle: At each
test point,
Average
trace no. n is calculated from
Average
trace
no. n – 1 and
Current
trace no. n according to:
)
(
)
(
1
)
1
(
1
)
(
c
n
n
Curr
c
n
Avg
c
c
n
Avg
>
+
−
−
=
Scalar quantities are averaged in analogy to
Average
traces. The formulas hold
for n = 1 where the average trace becomes equal to the current trace (statistics
off).
Calculation of
statistical quanti-
ties
In
Power
and
Modulation
measurements the statistical functions
Average, Mini-
mum, Maximum
and
Minimum/Maximum
are applied to a set of test points de-
pending on two independent parameters:
•
The time, i.e. the abscissa values t
i
, i ranging from 1 to the total number of
test points comprising the trace.
•
The burst number ranging from 1 to the number n of the current burst.
The result of the statistical operations depends on the parameter range consid-
ered and – in the case of statistics functions evaluated over several parameters
– on the order of evaluations. This is why the definition of statistical quantities
deserves some attention and is explained in the relevant sections in chapter 4.
Some particular examples are:
1. In
the
Power
menu, the quantity
Average Burst Power
denotes the average
power of the current burst. i.e. the arithmetical mean value of all test points t
i
located in the useful part of the burst (lower area 1 in the power template in
chapter 4).
2. In
the
Modulation
menu quantities such as the
Frequency Error
,
Phase Error
RMS, Phase Error Peak
etc. are first calculated for the current burst and
entered in the
Current
column of the output table
.
The results in the
Mini-
mum/Maximum
column correspond to the extreme value of the
Current
re-
sults calculated over all bursts measured. The results in the
Average
column
correspond to the average of the
Current
results calculated according to the
prescription in paragraph
Calculation of average quantities
above.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...